Expertise in scientific imaging

X-CITE XYLIS

X-Cite® XYLIS

Brilliance Across the Spectrum

X-Cite XYLIS is a tru​e arc lamp replacement for routine and advanced fluorescence imaging applications. It has the broadest spectrum available in a white light LED for fluorescence microscopy and rivals traditional arc lamps for brightness – making it ideal for both compound and stereomicroscopes. Incorporating our patented and award-winning LaserLED Hybrid Drive® technology to overcome the LED green gap, X-Cite XYLIS makes it possible to enjoy the benefits of LEDs without compromising on price, flexibility, or performance.

In addition to its powerful output and broad DAPI to Cy7 spectral range, X-Cite XYLIS offers the ultimate in flexibility – options are standard. Delivering light through a light guide alone or with a choice of more than a dozen microscope adaptors, X-Cite XYLIS can be installed on just about any new imaging system or used to retrofit the microscopes labs have depended on for years. Offering two models with a choice of UV wavelengths (365nm or 385nm), labs may choose the one that works with their preferred or existing DAPI filter sets. When it comes to control options, all are built-in to every X-Cite XYLIS – each system includes intuitive fingertip control with speedDIAL, hands-free operation with a foot pedal, and USB and TTL inputs for automated applications.

The X-Cite XYLIS provides an environmentally friendly replacement for lamp systems and has been tried and tested on demanding microscopy applications including Stereomicroscopy and FISH. The option to turn OFF the UV part of the spectrum will be welcomed by steromicroscope users who want to preserve the viability of their samples.

Specifications

Feauture & Benefits

FeaturesBenefits
Powerful LED TechnologyMercury-free, higher efficiency, stability, instant ON/OFF and long lifetime; reduce specimen exposure and scanning times to improve image quality and increase productivity
Broad Spectrum White Light ExcitationExcitation of commonly used fluorophores such as DAPI, CFP, GFP/FITC, YFP, TRITC/mCherry/Texas Red, Cy5 and Cy7
Two UV peak options: 365 or 385nmCompatible with any existing/preferred DAPI filter sets
“UV OFF” ModeEliminate UV exposure when not required, protect specimens and operators
Light Guide DeliveryFlexibility in light delivery as well as unit location
Multiple Control OptionsSimple operation through USB, TTL, speedDIAL or foot pedal (optional)
Optimized Thermal ManagementMaximum LED output, lifetime, stability and reliability
X-Cite Coupling Optics (optional)Efficient optical output and uniformity with the highest performing coupling optics available
Filter Recommendations  
FluorophoreFilter TypeSemrock FilterChroma Filter
DAPI (with XT710S, 365nm)SingleLED-DAPI-B49000
DAPI (with XT710L, 385nm)SingleLED-DAPI-B49028
CFPSingleLED-CFP-A49001
CFP/YFP FRETFRET multiple setFRET-CFP/YFP-C89002
FITC or GFPSingleLED-FITC-A49002
YFPSingleLED-YFP-A49003
TRITCSingleLED-TRITC-A49004
mCherrySingleLED-mCherry-A/mCherry-CmCherry-C49008
Cy5SingleLED-Cy5-A/Cy5-A-Basic49006
Cy7SingleLED-Cy7-A49007
DAPI/FITC/TexasRedTriple PinkelLED-DA/FI/TX‐3X‐A69302
DAPI/FITC/TRITC/Cy5Quad PinkelLED-DA/FI/TR/Cy5-4X‐B89400
CFP/YFP/mCherryTriple PinkelLED-CFP/YFP/mCherry-3X-A69308
GFP/mCherryDual PinkelGFP/HcRed‐2X‐A59222
DAPI/FITC/TRITC/Cy5/Cy7Penta PinkelLED-DA/FI/TR/Cy5/Cy7-5X-A
Fluorophores
FluorophoreWavelength (nm)
DAPI385
GFP/FITC475
TRITC/Cy3/RFP/mCherry560
MitoTracker® Red CMXROS580
Cy5640
Cy7735

 

Technical Specifications
 Model: XT720SModel: XT720L
Wavelength Range360-770nm380-770nm
LED Peaks (nm)365, 430, 475, 545, 650, 735385, 430, 475, 545, 650, 735
Input Power SupplyUniversal input 100-240VAC, 50/60Hz
Current3.9-1.6A
LED ON/OFF Response Times100 μs TTL / 1 ms USB
Control OptionsspeedDIAL
ON/OFF – TTL compatible
RS-232 commands (SDK available), USB
Foot Pedal (optional)
I/O ConnectionsBNC input
USB (B-type)
3.5mm stereo plug
Dimensions (W x H x D)120 mm x 260 mm x 260 mm (4.7″ x 10.3″ x 10.3″)
Weight6.1 kg (13.4 lbs)
CertificationsCE, RoHS
WarrantyLEDs – 25,000 hours or 3 years
All other X-Cite XYLIS components – 1 year, parts and labor (excluding LLG)
PatentsX-Cite XYLIS incorporates technology protected by patent US#9,239,133

Spectral Curve

X-Cite_XYLIS_Spectra

 

Control Options
Manual Control – speedDIALON/OFF
Speed sensitive intensity control
1 favorite intensity setting
UV OFF
Left and right-handed control option
USBON/OFF
Intensity control
Compatible with X-Cite 120PC, X-Cite exacte and
X-Cite 120LED drivers
TTLON/OFF
Foot Pedal (optional)ON/OFF
All X-Cite XYLIS models have input port, foot pedal sold separately

Resources